Fully Refurbished Pfeiffer Vacuum TPH 2200U Turbo Pump, Bottom Flange, DN 250 ISO, DN 63 ISO, F3 Oil, For TCP 600 Turbo Pump Controller, PMP02005 Overview: The Pfeiffer Vacuum TPH 2200U Turbo Pump, connecting cable and Electronic Drive Unit TCP 600 form, together with the appropriate backing pump, a system which is ready to operate. Switching on/off and turbo pump monitoring is executed via the TCP 600 Electronic Drive Unit. The TPH and the TPU Turbomolecular Pumps are of similar design and power. They differ in the type of high vacuum flange. - TPH: ISO-K Flange
- TPU: CF-F Flange
The pump rotor has bearings at both ends. Each ball bearing has its own circulatory lubrication system. The TCP controls pump running up to the rated rotation speed automatically. The pumps are thermally protected against excessive temperatures. If the temperatures of the bearings, the drive or pump casing are impermissibly high, the drive power is reduced to zero. Technical Data: Inlet Flange: DN 250 ISO-K Outlet Flange: DN 63 ISO-K Flow Rate (l/s): - N2: 2200
- He: 3200
- H2: 2800
Compression Ratio: - N2: 108
- He: 4 x 104
- H2: 2 x 103
Electronic Drive Unit: TCP 600 Rated Rotational Speed (1/min): 36000 Unit is fully rebuilt and comes with a FULL 6-MONTH WARRANTY! PN00171 AJVS Service and Repair Information:  Pfeiffer Vacuum TPH 2200U Repair A & J Vacuum Services provides repair service for all types of High Vacuum equipment. We perform Turbomolecular Pump repair, Turbo Pump Repair, Dry Pump repair, Vacuum Pump repair, Mass Spectrometer repair, Rotary Vane Pump repair, Vacuum Booster repair, and Leak detector repair. Our highly trained technicians have years of experience and guarantee the highest quality service. Due to our confidence in the quality parts and service we guarantee 6-12 month warranty for parts and labor. We operate at our own full service repair facility in Clifton, NJ. We have invested in state-of-the-art equipment including; complete line of decontamination, cleaning, drying, balancing, vibration analysis and software to minimize failures. Our in-house machine shop is capable of machining various components, prototypes and housing conversions as needed. We support all major manufacturers including Pfeiffer Vacuum, Alcatel / Adixen, Edwards Vacuum, Varian, Oerlikon / Leybod Vacuum, Vacuubrand, and Agilent Technologies. Repair Steps - Full incoming inspection and run up time check, vibration check, drive inspection.
- Full disassembly, decontamination and cleaning.
- Pump is balanced utilizing state-of-the-art equipment from the world leader in equipment balancing and analysis to reduce noise and vibrations and extend bearings life.
- Assembly and replacement of all necessary parts such as shafts, bushings, bearings, sleeves, O-rings, gaskets, seals, wicks, etc. Only parts that meet or surpass OEM specifications are used to ensure dependable highest quality service and long pump life.
- Final test and leak check. The pump in fully assembled condition is checked for vacuum seal. Leak test to 1 x 10-9 Mbar l/s. Vibrations are analyzed and test data is digitally recorded.
Every pump and instrument is extensively tested for 48 hours. During our final tests, we monitor total pressure, temperature, vibration and noise level utilizing newest testing technologies and techniques specified by OEM. Upon request, we can provide our customers with a pump down graph. Call us for your free repair estimate at 1-973-249-0854. For more information, please visit this products webpage. |